0102030405
China Suppliers Factory Temperature, Humidity, and Vibration Combined Test Chamber for High-Reliability Products
Technical Specifications
| Model | TAHPZ-150L | TAHPZ-225L | TAHPZ-408L | TAHPZ-800L | TAHPZ-1000L | |
| Control System | Balanced temperature and humidity control system | |||||
| Temp Range | -70℃ ~ 150℃ | |||||
| Humidity Range | 20℃ ~ 98%RH | |||||
| Temp Deviation | ±3.0℃ | |||||
| Humidity Deviation | ±3%R.H (≥75%R.H) / ±5%R.H (<75%R.H) | |||||
| Heating Up Time | -70℃ ~ 150℃ within 60 minutes | |||||
| Cooling Time | 20℃ ~ -70℃ within 75 minutes | |||||
| Inner Size (cm) | W | 50 | 50 | 60 | 100 | 100 |
| H | 60 | 75 | 85 | 100 | 100 | |
| D | 50 | 60 | 80 | 80 | 100 | |
| Inner Materials | SUS304 Advanced stainless steel | |||||
| Outer Materials | Cold rolled steel plate surface treatment by electrostatic spraying or SUS304 high-grade stainless steel | |||||
| Insulation Material | High density polyurethane foam and glass fiber | |||||
| Cooling System | Air cooled/water-cooled, single-stage/stacked, fully enclosed/semi enclosed compressors, environmentally friendly refrigerants (R404a/R23) | |||||
| Controller | LCD color touch screen, Chinese/English display, program, communication function | |||||
| Protector | Non fuse switch, compressor overpressure, overheating, overcurrent protection, overtemperature protection, fan overload protection, fuse, dry burning machine protector, water shortage protection switch, fault warning system, alarm | |||||
| Standard Accessory | Three-layer vacuum observation window, test hole, experimental stand, interior lighting, indicator light, mobile work wheel, humidification water tank, power cord, communication interface | |||||
| Vibration Table | Vertical and horizontal options can be selected according to user requirements | |||||
| Heater | High quality nickel alloy heater | |||||
| Blower | Centrifugal fan | |||||
| Power Supply | AC380±10% 50HZ | |||||
| Environmental Temp for Use | 5℃ ~ 35℃ | |||||
Key Highlights
-
1Synchronization Control System Achieve the linkage between the temperature and humidity controller and the vibration controller. Through a single software, the temperature and humidity curves as well as the vibration spectrum can be monitored simultaneously.
-
2Automatic Lifting System The box body realizes the automatic lifting function by means of guide rails or hydraulic devices.
-
3Anti-Condensation Technology During the alternating high and low temperatures, dry gas blowing is adopted to effectively prevent condensation.
-
4Isolation Vibration Treatment The box body and the vibration table base adopt an isolation design to prevent vibration transmission, ensuring the service life of the box body and the accuracy of the sensors.
Application
Compliance & Standards
- GB10589 Technical Conditions for Damp Heat Test Chamber
- GB2433.1-1989 Low Temperature Test Method
- GJB150.19-86 Vibration Test
FAQ
-
We are equipped with a special high-strength silicone flexible connection cover and a customized heat-insulating blind plate, which can effectively absorb high-frequency vibrations. Moreover, they remain uncracked and leak-free even at a -70℃ low temperature environment, ensuring stable and reliable sealing and shock resistance performance.
-
The equipment adopts an independent cooling air duct and insulation layer design, which can effectively prevent the heat generated by the vibration table motor from entering the test chamber.
-
Due to the strong vibration stress, an independent foundation (ground pit) or a large air cushion vibration isolation device is usually required. We can provide free foundation construction drawings.
-
The equipment uses environmentally friendly refrigerants R404a and R23, which comply with international environmental standards. The cooling system supports both air-cooled and water-cooled configurations with single-stage or stacked compressor options.
-
The chamber is equipped with comprehensive protection mechanisms including: non-fuse switch, compressor overpressure, overheating and overcurrent protection, overtemperature protection, fan overload protection, fuse, dry burning machine protector, water shortage protection switch, fault warning system, and alarm.
-
Yes. Both vertical and horizontal vibration table options are available and can be selected according to specific user requirements and test standards.