0102030405
High Pressure Accelerated Aging Chamber - China Suppliers & Factory for High Accelerated Life Testing Machines
Technical Specifications
| Equipment Model | TA-PCT-350 |
| Inner Tank Dimension | Φ350×400mm |
| Temperature Range | +105℃ ~ +132℃ |
| Humidity Range | Saturated Steam |
| Pressure Range | 1.0kg/cm² ~ 3.0kg/cm² (Gauge Pressure) |
| Circulation Mode | Natural Convection Circulation of Water Vapor |
| Pressurization Time | 0.00kg/cm² ~ 2.00kg/cm² within 45 minutes |
| Heating Time | Room Temperature to +132℃ within 35 minutes (Non-linear, No Load) |
| Controller | 7-inch Touch Screen + PLC |
Core Features
Dual-channel control: Supports dry and wet bulb temperature control, as well as saturated/unsaturated mode to prevent sample condensation.
Fast response: Can rapidly increase temperature and pressure to 130℃/85% RH/0.2MPa.
Automatic water supply: Equipped with automatic water replenishment and water quality filtration to protect the pipeline.
High-pressure safety: Circular pressure-bearing chamber, pressure interlock automatic door lock, multiple pressure relief protection.
Standards Compliance
JESD22-A110 / A118
IEC 60068-2-66
ASTM G154 / G15
Application
Semiconductor packaging
Optoelectronic industry
Precision electronics
Material research
Safety Features
Electronic lock: Automatically mechanically locked when pressure inside the box is too high.
Overpressure protection: Secondary mechanical pressure relief valve + electronic pressure relief alarm.
Anti-dry-burning: Automatically alarms and stops when there is insufficient water.
Leakage protection: Ensures the safety of operators.
Frequently Asked Questions (FAQ)
1. What are "saturated mode" and "unsaturated mode"? How to choose?
Saturation mode: The humidity is maintained at 100% RH, suitable for PCT high-pressure accelerated aging tests.
Non-saturation mode: The humidity can be adjusted (e.g., 85% RH), suitable for testing of electrically charged electronic components, avoiding the occurrence of condensation water that may cause false short circuits.
Non-saturation mode: The humidity can be adjusted (e.g., 85% RH), suitable for testing of electrically charged electronic components, avoiding the occurrence of condensation water that may cause false short circuits.
2. What are the differences between the HAST test and the traditional 85/85 test?
The traditional 85/85 test (at 85℃/85% RH) is an unpressurized test and takes approximately 1000 hours.
HAST enhances steam permeability through pressurization, achieving the same failure effect in only 96 to 168 hours, significantly shortening the research and development cycle.
HAST enhances steam permeability through pressurization, achieving the same failure effect in only 96 to 168 hours, significantly shortening the research and development cycle.
3. What are the requirements for water quality?
The HAST test has extremely strict requirements for water quality. Deionized water with an electrical conductivity of less than 5 μS/cm must be used, which can effectively prevent electrode corrosion and pipe scaling.
4. How does the safety door lock system operate during high-pressure cycles?
The chamber utilizes an automatic pressure-interlocking mechanism. When internal pressure is detected, the door remains mechanically locked and cannot be opened until internal chamber pressure safely returns to ambient atmospheric levels.
5. What international test standards does this test equipment satisfy?
This test chamber fully complies with recognized industry test standards, including JESD22-A110 / A118, IEC 60068-2-66, and ASTM G154 / G15 for highly accelerated stress and environmental corrosion testing.
6. How is internal condensation prevented from dripping onto test samples?
Through dual-channel dry and wet bulb temperature control alongside precise unsaturated test execution, the internal atmosphere maintains balanced equilibrium, effectively suppressing condensation buildup on tested component surfaces.