Explanation of the Humidity Function of the Constant Temperature and Humidity Chamber
We all know that a constant temperature and humidity chamber can be used for temperature testing. In addition to temperature, the name of the equipment also has the word 'constant humidity', so there is no doubt that it should be set up
Preparation has the function of humidity, so there are some precautions when using this function. It cannot be used at any time and casually. Therefore, in the following text, the editor will introduce the precautions for using humidity.
a. The humidity function cannot be turned on from the beginning. It can only be turned on after the temperature has stabilized and after one hour, otherwise it will cause significant temperature fluctuations.
b. D can turn the humidifier to a high level at once. If the humidity is too high, then turn it down again, or use a secondary humidity adjustment method. First, adjust the humidity to about 5 degrees lower than the required humidity. After waiting for the temperature, set it to the desired humidity.
c. There is a more convenient way to reduce humidity from high humidity to low humidity, which is to turn off the humidity, open the box door, wait for the humidity to drop to normal, and then turn on the humidity function again to save resources.
d. The temperature setting for the experiment is below 10 degrees, and the humidity function cannot be turned on because the temperature is too low, which will cause the evaporator to freeze and lose its function. So the above usage precautions are very important. Improper operation not only causes damage to the equipment, but also affects the accuracy of the experimental data. So when opening the humidity function of the constant temperature and humidity chamber, attention should be paid to the above factors.