Thermal Shock Chamber
Technical Specifications
1.Product Name |
Triple-Chamber thermal Shock Testing Machine |
2.Product Model |
TAS3-125LB |
3.Nominal Volume |
125L |
4.Effective Inner Box Dimensions |
500×500×500 W×H×D(mm) |
5.Outer Space: Approximately |
Approximately 1500×1900×1900 (W×H×D) (mm) |
6.Weight |
Approximately 1200kg |
7.Working Noise |
≤70db, measured 1.5 meters in front of the machine, 1.2 meters above the ground |
| |
1.Maximum Power |
Maximum power is approximately 19KW (subject to final design) |
2.Maximum Current |
Maximum operating current is 30A |
3.Cooling Water Requirements |
1. It is recommended to use a cooling water tower (one-to-one): ≥10T 2. Pipe Size: DN40, the length of the pipe layout is less than 20 meters. 3. Water Pressure: 0.3MPa~0.45MPa 4. Circulation Water Volume: ≥7.5m3/h 5. Water Temperature: 5~30℃ Note: This plan does not include the cost of the cooling water tower and water chiller and related piping. |
4.Power Supply Conditions and Power Source |
|
Product Features
Application
Compliance & Standards
FAQ
-
1.What are the differences between the three-box type cold and heat shock test chamber and the two-box type?
-
2.What are the requirements of the equipment for the installation environment?
Ground: The surface should be solid and level, capable of bearing the weight of the equipment (for large equipment, it can reach 1 to 2 tons).
Space: A distance of 60–100cm,should be reserved around the equipment for heat dissipation and maintenance operations.
Ventilation: The environment should be well-ventilated. It is recommended that the environmental temperature be within 5℃ to 30℃, to avoid a decrease in cooling efficiency or high-pressure alarms. -
3.How to choose between air cooling and water cooling?
Air-cooled: Installation is simple; just connect power and it can be used. It is suitable for small equipment or laboratories with low ambient temperatures.
Water-cooled: It has higher heat dissipation efficiency and is not limited by ambient temperature. It is suitable for high-power equipment and enclosed spaces, but it requires a cooling water tower and pipeline layout.